248mm
DIRECTIONS
1. To disinfect thermometer probe, wipe probe with or dip into isopropyl rubbing alcohol to a
maximum of ~ 0.25 inch from tip.
2. Press the On/Off Button next to the LCD. A tone will sound as the screen displays “
followed by “Lo”. The last recorded temperature will display, then the screen will again
display “Lo”. The thermometer is now in test mode and ready to obtain a temperature
reading.
M
”,
DIGITAL THERMOMETER, DISPOSABLE 60-SECOND
Model 1858
User Manual
WARNING: Read instructions thoroughly before using digital thermometer
*Note: Begin taking temperature within 15 seconds after thermometer enters test mode.
3. Position thermometer in desired location (mouth, rectum, or armpit) as directed below.
Do not use thermometer in the ear. Intended use is for oral, rectal, and axillary
(armpit) readings only.
a) Oral Use: To ensure accuracy, please keep mouth closed for 5 minutes before
testing. Place thermometer under tongue — recommended probe position is
indicated by “ √ ” shown in Figure 2. Readings are obtained in ~ 60 seconds.
Do not place thermometer near extreme heat - it may explode.
Note: Use of a probe cover may result in a 0.2°F discrepancy from actual temperature.
b) Rectal Use: Lubricate silver probe tip with petroleum jelly for easy insertion.
Gently insert probe ~ 1cm (less than 0.5 inch) into rectum. Readings are
obtained in ~ 60 seconds.
PLEASE READ CAREFULLY BEFORE USING
This digital thermometer provides a quick and accurate reading of an individual's body
temperature. Please read all instructions before using this thermometer.
Figure 2
c) Axillary (Armpit) Use: Wipe armpit dry. Place probe in armpit and keep arm pressed
firmly at side. Readings are obtained in ~ 90 seconds.
This thermometer is manufactured according to the system of EN12470-3:2000, ASTM
and complies with the requirements of IEC 60601-1:1990 +A1:1993+A2:1995,
IEC 60601-1-2:2007. The manufacturer is ISO 9001:2000 and ISO 13485:2003 certified.
4. The °F sign flashes throughout the testing process. When flashing stops, usually within
30~60 seconds, an alarm beeps for ~ 10 seconds and the measured reading simultaneously
displays.
CONTENTS
1 Thermometer, 1 User Manual, 1 Storage Case
5. To prolong battery life, press the On/Off Button to turn the unit off after testing is complete.
If no action is taken, the unit will automatically shut off after ~10 minutes.
On/Off
Button
PRODUCT DESCRIPTION
LCD
*Note: Normally the alarm beeps are “Bi-Bi-Bi-Bi”. The alarm will beep more rapidly when
temperature reaches 100°F or higher. When this occurs, the beeps will be
Probe
“Bi-Bi-Bi----- Bi-Bi-Bi------- Bi-Bi-Bi”. If room temperature is higher than 90°F
before testing, the LCD will display the actual room temperature instead of “Lo”.
Figure 1
PRECAUTIONS
* Do not bend or drop this thermometer. It is not shockproof.
* Do not boil thermometer probe to disinfect.
6-MONTH LIMITED WARRANTY
GF Health Products, Inc. warrants the HealthTeam® Model 1858 Thermometer against manufacturer's defects
for a period of six months. If a product is deemed to be under warranty, GF Health Products, Inc. shall provide,
at its option, (1) replacement of any defective part or product or (2) a credit of the original selling price made
to GF Health Products, Inc.'s initial customer. The warranty does not include any labor charges incurred in
replacement part(s) installation or any associated freight or shipping charges to GF Health Products, Inc.
* Do not store in direct sunlight or in areas of extreme heat.
* Do not disassemble. The battery is not replaceable.
*
*
--- Direct Volts D.C.
--- Type B Equipment.
GF Health Products, Inc. is not responsible for typographical errors. Packaging, warranties, products and
specifications are subject to change without notice.
SPECIFICATIONS
Graham-Field and HealthTeam are registered trademarks of GF Health Products, Inc.
Range:
90°F - 109°F
Accuracy:
Display:
Memory:
±.2°F > 95.9°F and < 107.6°F; ±.4°F < 95.9°F or > 107.6°F
Liquid crystal display, XXX.X digit display
For storing the last measured value
Battery:
One 1.5 V DC battery (battery is not replaceable)
Power consumption:
Device (battery) life:
0.15 milliwatts in measurement mode
~ 200 hours
5 inches long x .71 inch wide x .31 inch high
~ 0.02 lb / 0.35 oz / 10 g
Dimensions:
Weight:
GF Health Products, Inc.
2935 Northeast Parkway
Atlanta, GA 30360
Tel: 800-347-5678
Fax: 800-721-0601
Operating environment:
41°F - 104°F, < 85% RH
14°F - 131°F, < 95% RH
Storage and transportation
environment:
1858-INS-LAB-RevA09
|